EasyManuals Logo

Mindray A7 Service Manual

Mindray A7
482 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #223 background imageLoading...
Page #223 background image
A7â„¢ Service Manual 046-006272-00 4 - 11
Calibration System Calibration
FIGURE 4-11
b. Zero Mode Settings: Press the Setup button, select Setting->ENTER->Zero Mode->Manual-
>BACK->BACK.
FIGURE 4-12
c. Serial Mode Settings: Press the Setup button, select Setting->System->Enter->Serial Mode -
>OTIS Ctrl->BACK->BACK.
FIGURE 4-13
d. After setting up the calibration device, the calibration enters the serial mode screen shown
below.

Table of Contents

Other manuals for Mindray A7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A7 and is the answer not in the manual?

Mindray A7 Specifications

General IconGeneral
TypeAnesthesia Workstation
BrandMindray
ModelA7
Display15-inch color touchscreen
Oxygen FlushYes
Gas InletsO2, N2O, Air
Ventilation ModesSIMV, PSV
Breathing SystemIntegrated
Gas MonitoringO2, CO2, N2O
Monitoring ParametersFiO2
CO2 MonitoringMainstream or Sidestream
Power Supply100-240 V, 50/60 Hz

Related product manuals