A7™ Service Manual 046-006272-00 4 - 19
Calibration System Calibration
4.3.2.3 Manual Calibration
Follow these steps to calibrate the flow sensors.
1. Enter Standby.
2. Select Setup-> Service-> Calibration-> Flow Sensors to access the screen as shown.
FIGURE 4-27
3. Select Calibrate Manually button to open the menu shown below.