EasyManua.ls Logo

Mindray A7

Mindray A7
482 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
System Calibration Calibration
4 - 24 046-006272-00 A7™ Service Manual
FIGURE 4-36
12.Select Cancel and the screen shown below is displayed.
FIGURE 4-37
NOTE: After flow calibration, check the accuracy of flow sensors by referring to
Section 3.8.7 Check the Flow Sensor Accuracy.
NOTE: In case of calibration failure, first fix the problem and then perform flow
calibration again.

Table of Contents

Other manuals for Mindray A7

Related product manuals