EasyManua.ls Logo

Mindray A7

Mindray A7
482 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
System Calibration Calibration
4 - 32 046-006272-00 A7™ Service Manual
FIGURE 4-39
4. Connect the calibration device with the anesthesia machine using communication cable by
referring to step 4. of ‘Auto Calibration’’ on page 4-8
5. Press the Next button to open the menu shown below.
FIGURE 4-40
A four-way device is required to connect the sampling lines for pressure calibration. The following
pictures show the four-way device, connectors on the calibration device and ventilator control board
involved for pressure calibration.

Table of Contents

Other manuals for Mindray A7

Related product manuals