System Calibration Calibration
4 - 36 046-006272-00 A7™ Service Manual
FIGURE 4-46
11. The screen shown below is displayed if the ongoing calibration is canceled. Select Try Again to
do the calibration again. Select Done to exit the calibration screen.
FIGURE 4-47
4.3.3.1.2 Manual Calibration
Follow these steps to calibrate pressure sensors and the PEEP valve.
1. Make sure that the anesthesia machine is in standby mode.
2. Select Setup-> Service-> Calibration-> Pressure Sensors to access the screen shown below.