A7â„¢ Service Manual 046-006272-00 6 - 7
Repair and Disassembly Disassemble the Assemblies
6.2.1.6 Remove the Electronic Flowmeter
1. Remove the EFCS module.
2. Disconnect the cables from the electronic flowmeter.
3. Unscrew the four screws to remove the O2, N2O and AIR electronic flowmeters. Take care not to
drop the sealing rings below the electronic flowmeters.
FIGURE 6-7