Disassemble the Assemblies Repair and Disassembly
6 - 26 046-006272-00 A7â„¢ Service Manual
Remove the Auxiliary Outlet Assembly
1. Remove the rear panel assembly of the hardware box.
2. Unscrew one nut and disconnect the cable from the breaker.
FIGURE 6-49
3. Unscrew the two screws on each auxiliary outlet to remove the auxiliary outlets. The A5 has four
(4) auxiliary outlets, as shown in the photograph below. The A3 has three (3) auxiliary outlets
(not shown).
FIGURE 6-50