EasyManua.ls Logo

Mindray BC-2800 - Page 242

Mindray BC-2800
286 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Troubleshooting Your Analyzer
11-6
RBC count time and the actual RBC
count time is less than 2 seconds, the
error has been removed;
4. If not, enter the “Service
Maintenance” screen and do the
probe cleanser cleaning procedure as
instructed by Chapter 10.3.6.;
5. Enter the “Setup Settings
Count” screen and record the RBC
count time. Then enter the “Service
Self-test” screen and test the actual
RBC count time as instructed by
Chapter 10.5.1.;
6. If the difference between the reference
RBC count time and the actual RBC
count time is less than 2 seconds, the
error has been removed;
7. If the difference is still greater than 2
seconds but consistent, enter the
Setup Settings Count” and
reset the RBC count time. Then enter
the “Service Self-test” screen and
test the actual RBC count time as
instructed by Chapter 10.5.1 to
confirm the difference is less than 2
seconds;
8. If the problem remains, contact
Mindray customer service department
or your local distributor.
RBC bubbles
1. Diluent or rinse running
out;
2. Loose tube connections;
3. Inappropriate RBC
count time setting.
1. Check if the diluent or rinse has run
out. If so, change a new container of
diluent or rinse as instructed in
Chapter 4.4.1;
2. Check the connection of the diluent
and rinse pickup tube. If necessary,
reconnect and tighten them as
instructed by Chapter 4.4.1;
3. If the problem remains, adjust the RBC
count time as instructed by Chapter
5.3.5;
4. If the problem remains, contact

Table of Contents

Other manuals for Mindray BC-2800

Related product manuals