5-8 Function and Performance Checking Method
2. M mode
In M mode scanning, the image parameter area in the upper left corner of the screen displays
the real-time parameter values as follows:
Parameter
F
D
G
V
DR
Meaning
Frequency
Depth
M Gain
M speed
M Dynamic Range
Parameters that can be adjusted to optimize the M Mode image are indicated in the following.
Control Panel
Gain, Depth, TGC, Focus position
Touch Screen
Speed, Display Format, Gray Map, Dynamic Range, Tint Map, M Soften, Edge
Enhance
3. Color mode
In Color mode scanning, the image parameter area in the upper left corner of the screen
displays the real-time parameter values as follows:
Parameter
F
G
PRF
WF
Meaning
Frequency
Color Gain
Pulse Repetition Frequency (PRF)
Color Wall Filter
Parameters that can be adjusted to optimize the Color Mode image are indicated in the
following.
Control
Panel
Gain, Depth, iTouch, Scale
Touch
Screen
Invert, Baseline, B/C Wide, Dual Live, Image quality, Flow State, Priority, Packet Size,
Map, Wall Filter, Line Density, Smooth, Persistence, Velocity Tag, steer
4. Power mode
In Power mode scanning, the image parameter area in the upper left corner of the screen
displays the real-time parameter values as follows:
Display
F
G
WF
PRF
Parameter
Frequency
Power
Gain
Power Wall
Filter
Pulse
Repetition
Frequency
Echo Boost
B image touch screen-[Echo
Boost].
[Echo Boost] is enabled when it is on in B mode. (Highlighted) the
system is in “Echo Boost” status.

Table of Contents