5-98  Image Optimization 
Operation  Tap [Invert] to enable the function (the soft key is highlighted). 
 
iLayering 
Description  To increase the layer display in elasto images. 
Operation  Tap [iLayering] to enable the function. 
 
Filtering 
Description  To filter the noise of the elasto image. 
Operation  Tap  [Filter] on the touchpad. 
The higher the level is, the smaller the noise of the elasto image becomes, and the 
clearer the edge of the field target appears. 
Impacts  The system restarts scanning B image and E image after the filtering is completed. 
 
Map Position 
Description  This feature is used to adjust the up/down position of the map.   
Operations  Tap [Map Position] on the touch screen to adjust the parameter.   
 
Select/Bad 
Description  It is used to remove the desired frame from stress curve. The removed frame will not 
be calculated in STE indices. 
Operations 
Rotate the knob under [Select/Bad] to select the desired frame, and press the knob 
to set the desired frame to “On”. 
 
5.14.2.3  Measuring 
Press <Caliper> to enter general measurements. You can perform the measurements of Elastography, 
Elas. Ratio, Directional Ratio, Elas. Hist and etc. 
Press <Measure> to enter application measurement. You can perform measurements on Isthmus and 
Mass, etc.   
Refer to Operator’s Manual (Advanced Volume) for details. 
5.14.2.4  Cine Review 
Press <Freeze> to freeze the image and enter cine review state. 
Or, click the cine to enter the cine review state. 
5.14.3  STQ Imaging (Sound Touch Quantification) 
The STQ imaging is an option. Probes support this function:   
 
5.14.3.1  Basic Procedures for the STQ Imaging 
1.  Select a proper probe. Perform 2D scan to locate the lesion.