12-4  Setup 
Type  Item  Description 
Screen Saver 
Enable Screensaver 
Select the different saver methods to the system. 
After enabling the screen saver, check “Mindray” to select 
the image from the system. Or click [Preview] to select the 
image on your own. 
To set the waiting time before the system enters dormancy 
status in the drop-down list beside “Wait”.   
The system enters screen saver automatically if the 
system waiting time exceeds the screen saver already set. 
Enable Standby 
The system enters standby status automatically if the 
system waiting time exceeds the screen saver already set 
and standby time. 
iVocal 
Enable Voice 
Response 
To enable the voice response function. 
Automatically shut 
down waiting time   
To automatically disable the voice response function after 
the setting time.   
12.1.3  Image Preset 
Open the page via [Setup]  [System] [Image]. 
 
 
 
 
Reset Config. 
 
 
Probe 
To set the default probe model for the system from the 
drop-down list. 
The default parameters are applied to the new probe if 
checking Use the default setting when start a new exam.   
Image size 
Set the standard for saving the image or using the 
digital/graph printer. 
Default Elasto  Set the default elastography mode. 
ART Flow 
Max transmit time 
Set the maximum duration time. Enter Color mode and 
enable ART Flow; the penetrability of color image is 
enhanced in this time period. 
Max interval time 
Set the maximum interval time. Enter Color mode and 
enable ART Flow; the ART Flow is enabled again until the 
Parameter 
Steer 
To set the steer mode in B + Color + PW/CW imaging 
mode. 
C&(PW/CW): select to adjust the sample volume in color 
mode and sample line in PW or CW mode together. 
C/(PW/CW): select to adjust the sample volume in color 
mode and sample line in PW or CW mode separately. 
Auto Invert 
The spectrum can automatically invert when the color flow 
is steered to a certain angle, thus accommodating the 
operator’s wish to distinguish the flow direction. 
iScape Ruler Display 
To set whether to display the iScape ruler in iScape 
imaging mode.