System Preparation 3-3
WARNING:
1.
Be sure to connect the equipotential wire before inserting the power
plug into the receptacle; be sure to pull out the power plug from the
receptacle before disconnecting the equipotential wire; otherwise
electric shock may result.
2.
When you connect another device to this system, you should use
the equipotential wire to connect each of equipotential terminals;
otherwise electric shock may result.
3.
Connect the earth cable before turning ON the system. Disconnect
the earth cable after turning OFF the system. Otherwise, electric
shock may result.
4.
DO NOT connect this system to outlets with the same circuit
breakers and fuses that control the current to devices such as life-
support systems. If this system malfunctions and generates
overcurrent, or when there is an instantaneous current at power ON,
the circuit breakers and fuses of the buildings supply circuit may be
tripped.
3.2.4 Powering on the System
CAUTION:
To ensure safe and effective system operation, you must perform daily
maintenance and checks. If the system begins to function improperly
immediately stop scanning. If the system continues to function improperly
fully shut down the system and contact Mindray Customer Service Department
or sales representative. If you use the system in a persistent improperly
functioning state you may harm the patient or damage the equipment.
Checking before Power ON
Check the system before the system is powered on:
No. Check Item
1
The temperature, relative humidity and atmospheric pressure shall meet the
requirements of operating conditions. For details, please refer to 2.3.2 Environmental
Conditions
2 There shall be no condensation.
3
There shall be no distortion, damage or dirt on the system and peripheral devices.
If any dirt is found, cleaning shall be performed.
4 There shall be loose screws on the monitor or control panel.
5
There shall be no cable damage (e.g. power cord). Maintaining secure connections to
the system at all times.
6
The probes and probe cables shall be free of damage or stains.
See Chapter 13 Probes and Biopsy for details of probe cleaning and disinfection.
7
No miscellaneous odds and ends are allowed to be attached or affixed to the control
panel.
8
Ensure that all connections are free from damage and remain clear of foreign object
blockages.
There shall be no obstacles around the system and its air vent.
9 Probe cleaning and disinfection.
10 The overall scanning environment and field must be clean.