3-4 System Preparation
No. Check Item
11
The locking mechanism of casters can work normally.
To check the system after it is turned on
Before getting the system started, check if the system is plugged in verify that the system circuit
breaker is at On”. After the power indicator on the touch panel becomes green, press the power
button to power on the system. The system image appears.
To check the system after the system is turned on:
No. Check Item
1 There shall no unusual sounds or smells indicating possible overheating.
2 There shall be no persistently displayed system error message.
3 There shall no evident excessive noise, discontinuous, absent or black artifacts in
the B Mode image.
4
Check if there is abnormal heat on the surface of the probe during an ultrasound
procedure. If you use a probe giving off excessive heat, it may burn the patient.
5 The control panel keys and knobs are fully functional.
6 The touch screen and the main monitor screens display normally depending on
the system modes and image status.
7 The exam date and time are the same as the system date and time, and are
displayed correctly.
WARNING:
1. The patient may get burnt if using an excessive heated probe.
2. The system is defective if any malfunctioning happens. In this case,
shut down the system immediately and contact Mindray Customer
Service Department or sales representative.
NOTE:
If there is over loading to the system, the breaker will switch into OFF to discontinue the
power supply. If the breaker cannot be set as On, or returns to OFF after being switched on,
please disconnect the power cables, and contact your sales representative.
3.2.5 Powering off the System
You need to follow the correct procedures to power off the system. In addition, after you upgrade the
software or when the system is down, you need to power off and restart it.
To power off your system normally
Press the power button on left of the touch panel to see the option:
Shut down: To power off the system normally.
Standby: To enter standby status.
Cancel: To cancel the operation.
If you will not use the system for a long period of time, you should turn off the circuit breaker and
disconnect the power to all other peripheral devices.
NOTE: Do not turn off the breaker while the power indicator is on. Otherwise, the data may lose
or the system software may be damaged.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Recho R9 and is the answer not in the manual?

Mindray Recho R9 Specifications

General IconGeneral
BrandMindray
ModelRecho R9
CategoryMedical Equipment
LanguageEnglish

Related product manuals