EasyManuals Logo

Mindray Recho R9 User Manual

Mindray Recho R9
477 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #81 background imageLoading...
Page #81 background image
Image Optimization 5-7
Impacts The FOV position/range is available for the convex, linear and phased probes.
Steer is available only for linear probes.
Line Density
Description The function determines the quality and information of the image.
Operation Tap [Line Density] on the touch screen to adjust the parameters.
Levels of line density: UH/H/M/L.
Impacts
The higher the line density is, the higher the resolution becomes.
Dynamic Range
Description Adjusts contrast resolution of an image, compresses or expands gray display range.
The real-time dynamic range value is displayed on the image parameter area in the
upper left corner of the screen.
Operation
Rotate the knob below [Dyn Ra.] to adjust the parameters.
Rotate the knob clockwise to increase the value; rotate the knob counterclockwise to
decrease the value.
Impacts
The more the dynamic range, the more specified the information, and the lower the
contrast with more noise.
Smooth
Description
This feature is used to reject the noise and smooth the image.
Operation
Rotate the knob under [Smooth] to adjust the parameter.
The bigger the value is, the higher the smooth becomes.
iClear
Description
This function is used to increase image profile, so as to distinguish the image
boundary.
Operation Rotate the knob under [iClear] to adjust the parameters.
Off represents iClear is disabled, and the bigger the value is the stronger the effect
becomes.
iClear
+
Description
This function is the advanced version of iClear. It can better enhance image quality,
reject noise, and provide better contrast resolution.
Operation Rotate the knob under [iClear
+
] on the touch screen to adjust the parameters.
Off represents iClear
+
is disabled, and the bigger the value is, the stronger the effect
becomes.
Impacts iClear
+
is an option. If iClear
+
is configured, iClear will be replaced by iClear
+
.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Recho R9 and is the answer not in the manual?

Mindray Recho R9 Specifications

General IconGeneral
BrandMindray
ModelRecho R9
CategoryMedical Equipment
LanguageEnglish

Related product manuals