5-10 Image Optimization
Operation Press <iTouch> on the control panel to enter the iTouch status, the symbol of which will
be displayed in the image parameter area of the screen.
Long press <iTouch> to exit.
Tap [iTouch] on the touch screen to adjust the parameters.
HScale
Display or hide the width scale (horizontal scale).
The scale of the horizontal scale is the same as that of vertical scale (depth), they
change together in zoom mode, or when the number of the image window changes.
The HScale will be inverted when image is turned upwards/downwards.
Operation Click [H Scale] on the touch screen to display or hide the scale.
Dual Live
Display different image effects of one probe for a better observation.
Tap [Dual Live] on the touch screen to enable/disable the function, and dual-split
window of images are displayed.
Two pages of adjustable parameters are displayed on the touch screen as well; where,
shared parameters and left window parameters are displayed in the B (L) page, while
right window parameters are displayed in the B(R) page.
In the image parameter area on the upper left corner of the main screen, parameters
of the both windows are displayed.
Impacts It supports the magnification of the image.
LGC
Description Adjust the gain along the scan line to improve the lateral resolution of the image.
Operation Tap [LGC] on the touch screen to enter the adjusting dialogue box.
The 8 rods displayed on the touch screen indicate the corresponding image areas on
the main screen.
Tap the slider to adjust the gain, move downwards to decrease the gain, while move
upwards to increase the gain.
In addition, the system provides several preset parameters for imaging.
Echo Boost
Description
The contrast is increased and the noise is decreased with the clear boundary after
generating the function.
Operation
Tap [Echo Boost] on the touch screen. Echo Boost value appears on the parameter
area.
Use phased probe to activate the function in cardiac mode.