52
Signum 3 Ex
Neutral Measurement Application Z
nM
Parameter for Saving Weight Values
The reference weight is saved when the
scale has stabilized�
“Stability” is defined as the point at
which the fluctuation of a measured
value lies within a defined tolerance
range� The narrower the tolerance
range, the more stable the platform is
at “stability”�
In Setup menu:
appl1
:
Neutr.M:3.11.
You can dene whether the value is
saved when “standard stability” is
reached, or only at “increased stability”
(narrower tolerance range)� If you select
“increased stability”, the reference
weight saved will be more accurate and
the results more reproducible, but the
response time of the weighing platform
might be longer�
Accuracy Level for Calculation of
Reference Value
The resolution applied for calculating
the reference weight is defined in Setup
under:
appl1
:
Neutr.M: 3.9.
The resolution for calculating the
reference weight is increased if “+1
decimal place,” “+2 decimal places” or
“Internal resolution” is selected� With
the “+1 decimal place” setting, the net
value is determined to one additional
decimal place (i�e�, display accuracy
x 10); “+2 decimal places” increases
display accuracy x 100, and so on up to
the maximum resolution available�
Decimal Places in Displayed Result
In neutral measurement, not only whole
numbers but also decimal numbers
(for example, 1�25 o electrical cabling)
can be displayed� The number of
decimal places displayed in neutral
measurement is congured in Setup
under:
appl1
:
Neutr.M: 3.10.
Minimum Load
The minimum load required for
initialization of the weighing platform
is configured in Setup under:
appl1
:
Neutr.M: 3.6.
Once the limit is exceeded by the load,
initialization can begin�
If the load on platform is too light, the
following will occur when you try to
save a value:
– Error code Inf 29 appears,
– A warning signal is emitted (double-
beep)
– The weighing platform is not initialized
– The preset calculation factor is saved
The minimum load required for
automatic taring of the container
weight on the platform (rst weight)
iscongured in Setup under:
appl
: m.weigh: 3.5.
You can choose from the following
10levels for this setting:
1 digit
2 digits
5 digits
10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
The “digits” here refer to the scale
intervals in the connected weighing
platform� If the interval of the
connected platform is 1 g, for example,
and 1000 digits are required, you
must place at least 1000 g (= 1000
intervals) on the weighing platform for
initialization�
Neutral Measurement Application Z
nM