71
Signum 3 Ex
Net-total Formulation Application R Net-total Formulation Application R
Preparation
§ Turn scale on: Press the e key
§ While all segments are lit, press the
)key
§ Select Setup: Press the k key
repeatedly until appl is displayed
§ Conrm appl: Press the ) key
§ Selecting the Net-total Formulation
Application:
Press the k key repeatedly and
conrm with the ) key
Net-total Formulation Application
Parameters
3�6� Minimum load for autosave
3�6�1* 1 digit
3�6�2 2 digits
3�6�3 5 digits
3�6�4 10 digits
3�6�5 20 digits
3�6�6 50 digits
3�6�7 100 digits
3�6�8 200 digits
3�6�9 500 digits
3�6�10 1000 digits
3�17� Individual/Component printout
when saved
3�17�1 Automatic printing o
3�17�2* Print the entire standard
print conguration
every time with the
O key
3�17�3* Print the entire standard
print conguration once
with the O key
* = Factory setting
§ Save the settings with the ) key and
exit Setup: Press the ( key several
times�
Minimum Load
The minimum amount that a
component must weigh before it
can besaved in net-total memory is
congured in Setup under: appl 3
Net Tot: 3.6.
Once the limit is exceeded by the load,
the value can be saved� If the load on
platform is too light, the following will
occur when you try to save a value:
Error code Inf 29 appears,
The weight is not saved
The minimum load required for
automatic taring of the container
weight on the platform (rst weight)
iscongured in Setup under:
Appl: m.weigh: 3.5.
You can choose from the following
10levels for this setting:
1 digit
2 digits
5 digits
10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
The digits” here refer to the scale
intervals in the connected weighing
platform� If the interval of the
connected platform is 1 g, for example,
and 1000 digits are required, you must
place at least 1000 g (= 1000 intervals)
on the weighing platform for saving�