Chapter 5 Solutions for Errors 5-
5
Unwanted data is
continually received.
Communication error. Check the
connector and retry. If the error is
not resolved, replace the board. If
the error occurs frequently,
remove dust from the board
(KVME-080) and retry.
If the error is not resolved, check
the cable and the contact the C/S
engineer.
ACK not received.
(Port 0 – Serial Shell)
Unidentifiable error. Ignore and
continue. If the error is not
resolved, check again after power
Off/On. If the same error occurs
repeatedly, contact the C/S
engineer.
Ignore and continue. If the error is
not resolved, check damage of the
line from the Board (KVME-080)
port.
STX not received.
(Port 0 – Serial Shell)
Unidentifiable error. Ignore and
continue. If the error is not
resolved, check again after power
Off/On. If the same error occurs
repeatedly, contact the C/S
engineer.
Check connection of the cable and
board port. If the error is not
resolved, contact the C/S
engineer.
Check connection of the cable and
board port. If the error is not
resolved, contact the C/S
engineer.
Unidentifiable error. Ignore and
continue. If the error is not
resolved, check again after power
Off/On. If the same error occurs
repeatedly, contact the C/S
engineer.
STX error
(Port 1 - T-
Box)
Check that T-Box power is on and the
connection of the T-Box cable and
Board port1. If the error is not
resolved, replace the T-box. If the error
occurs repeatedly, it is probably a
defect of the line or the board (KVME-
080) itself. Replace the detective parts.
Call the C/S Engineer from the T-Box.
STX error
(Port 2 - Front VSC)
Not applicable to Mx Series