60 | P a g e
mechanical switch. Replace if inoperable.
Reinstall components in reverse order.
4. (REF 8.2.4.2)
My hand rinse or end rinse are
not working
Hand spray hose kinked
Un-kink hand spray hose
Check in lower skirting for interference
Mixing valve “OFF”
Turn mixing valve “ON”
Gate valve for end rinse
“OFF”
Turn gate valve “ON
Water system plugged
Check Y Strainers for debris.
Check for kinked supply lines.
Check backflow prevention devices (REF 8.2.5)
Clean end rinse orifice with small pick to dislodge
debris.
Station is leaking water
Leak at facility connection
Check strainer drain plug
Check flex lines from strainer.
Check drain line connection.
Check P-trap to facility connection
Leak internal to station
Shut off all mixing valves and turn off hot/cold
separately to isolate leak circuit (REF 9.3)
Service water lines & connections (REF 9.3)
LIGHTING
No lights will come on
G.F.C.I. has been tripped
Reset G.F.C.I. (station power must be “ON”)
Individual 7” overhead light not
working
Bad Light driver or loose
connection at driver
Check for loose connection at driver, replace if
necessary
LED failure
Replace LED light
Surface light not working
Surface light not turned “ON”
in touch screen
Turn surface light “ON” in touch screen Lighting
menu screen
Power supply issue
Check connections at harness for LED
Check connections at distribution blocks/DIN
Check DC power supply function/output
LED failure
Replace fixture
Shelf-mounted task light not
working
Power switch in light base
not turned “ON”
Turn power switch in light base “ON”