Pressure measurement
Pressure (4 values) is measured and saved to the selected measurement
name. The actual measured value is displayed in the middle of the dis-
play. The 4 measurement names can be changed as desired.
The hose on the draft + connector must be connected for draft meas-
urements.
The second hose on the Delta P- connector must be connected for dif-
ferential measurement.
► Go to the Measurement menu.
► Select Pressure measurement.
► Press OK.
The Menu Pressure measurement appears.
► If necessary, press the Menu Key to change the name.
A selection list appears.
► Select Modify name.
A window appears.
► Adapt the name to your needs.
► Press OK.
The changed name appears in menu Pressure measurement.
► If necessary, press F1 to accept the current value.
► If necessary, press F2 to take a zero point.