USER MANUAL OPTIMA 7
MRU GmbH, D-74172 Neckarsulm 42 / 78
Temporary buffer
The analyzer gives the possibility to set the momentary values into a
temporary buffer during effecting and continuing the measurement.
► Later on, the values can be brought back from the temporary buffer
to the measuring window in order to print them out or / and to save
them.
Store measured values in the buffer
During an actual measurement you can set the actual values into the
temporary buffer
► Start a measurement.
See also Chapter 7.1 Selecting the measuring program, Page 35.
► Press the Menu Key.
A selection list appears.
► Select values to clipboard.
The measured values are stored in the temporary buffer.
If offered, you can store measured values directly from the function key
bar in the temporary buffer.