USER MANUAL OPTIMA 7
MRU GmbH, D-74172 Neckarsulm 76 / 78
Troubleshooting condensate separator
1. Effect
2. Cause
3. Solution
Dirt and / or humidity in-
side the device
No filter effects
Sensor failure
Pump failure
Fine filters are wet and /
or dirty.
Check filters more
often
Renew them if nec-
essary (white = OK)
Brown-black = re-
newal
Wrong measuring values
Cover, intermediary unit,
plexiglass tube and lock-
ing pieces are not tightly
fixed respectively screwed
Check tightness
with every filter
change.
Further options
Prefilter for high dust concentrations
If necessary, this pre-filter is inserted into the hose line directly after the
probe handle.
Filter Tablets:
# 52798 50 Filter Tablets