To Previous Page IconTo Previous Page
Loading...
MSA AUER
MSA
Operation of the System
SUPREMATouch
45
US
Status [Status of this data cell]
Field type: Display
This field displays the status of the cell used for data saving. If this status is “protected”, then no
data can be changed and the following input fields are simply display fields.
2 x Name [Head name in English [upper] and local language [lower]
Field type: text
In these fields, the name of the remote sensing head is displayed in both supported languages.
The remote sensing head can be selected as “Sensor type” in the setting of the measuring points
through these names.
The user can freely define the names in changeable remote sensing heads. They must be unique,
that is, no name should be given twice. If a name is given for only one language, the same name
can also be used for the other language while saving it.
UAmin [Limit for “suppressed”]
Field type: Number [integers]
Adjusting range: 240...350
This field displays the minimum signal UA for the status “suppressed”. Measuring values below
this limit value are displayed as “suppressed”. If field is empty [tap “Clear” in the number input win-
dow], then this status will not be tested.
This value can only be adjusted for changeable remote sensing heads.
UAidle [Limit for signal failure]
Field type: Number [integers]
Adjusting range: 50…UAmin
This entry displays the minimum signal UA for the status “Signal failure”. Measuring values below
this limit value are displayed as “Signal failure”.
This value can only be adjusted changeable remote sensing heads.
UAover [Limit for “overflow”]
Field type: Number [integers]
Adjusting range: 2000…2200
This value defines the sensor signal UA of the remote sensing head for the measuring range over-
flow display. Measuring values above this value are displayed as “overflow”.
This value can only be adjusted for changeable remote sensing heads.
Tsupp [Warm-up period]
Field type: Number [integers]
Adjusting range: 10…300