EasyManua.ls Logo

MTE HYDROCAL1005 - Page 51

MTE HYDROCAL1005
113 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
HYDROCAL 1005 Manual for Installation and Operation Page 49/111
After a measuring variable has been selected and opened with for configuration, the following
functions are available:
Selection of different input possibilities for the alarm mode (mode), Action (action1..2)
and choice of the output (clamp1..2)
Activation of input over alphanumeric keyboard for the values: Alarm limit (Level1..2),
Delay (Delay1..2)
Scroll up
Scroll down
Scroll up / down
Cancel alarm
Store configuration of alarm and exit
Exit without storage
Following alarm settings can be chosen by pressing the button :
High: An alarm is triggered when Level 1 is exceeded.
High-High: An alarm in 2 steps is triggered. The first alarm is triggered when Level 1 is ex-
ceeded; the second alarm is triggered after exceeding Level 2.
Low: An alarm is triggered when a value falls below the limit defined in Level 1.
Low-Low: An alarm in 2 steps is triggered. Step 1 is alarm is triggered when the measured value
falls below the limit defined in Level 1, the second alarm is triggered below Level 2 when the
measured value falls below the limit defined in Level 2.
Low-high: An alarm in 2 steps is triggered. Step 1 is triggered when the measured value falls be-
low the limit defined in Level 1, the second alarm is triggered after exceeding Level 2.
Range: An alarm is triggered when the measured value is between the limits defined in Levels 1
and 2.
The alarm limits are edited by arrow keys and confirmed by the button and the alphanumeric
keyboard. The quantity is defined by the unit. This is “ppm” for gas concentration and „°C” for tem-
peratures and for external sensors the unit defined for the respective sensor (e.g. Volt, Ampere ...).
The alarm limits must be within the corresponding measuring ranges. This is 0...2000 ppm for gases,
0...+100°C for temperatures and for external sensors the limit is defined for the respective sensor.
Delays (Delay1...2) can be programmed to suppress alarms if the limits are exceeded only for a
short time. Input of delay time is made in seconds or by a button in days hours min - seconds.
For minus the button and for a double point the button are used. After confirmation by the
button, the value will be converted and displayed in days hours min - seconds.

Table of Contents

Related product manuals