Calibrating a Force Sensor
Calibration
136
and the software will calculate the Pre-Amp and Post-Amp gain
values automatically.
When you are calibrating an DC conditioner, use the Post-Amp gain
control to increase gain. If more gain is needed, you must disable
hydraulics and increase the Pre-Amp gain. You can then turn on
hydraulics and continue to adjust the Post-Amp gain.
1. Select Gain/Delta-K for Cal Type.
Stand-alone— Path: Setup > Force > Calibration > Cal Type >
Gain/Delta-K
Automated— Select Gain/Delta-K for Cal Type on the
Calibration tab of the Inputs panel.
2. Exercise the force standard.
Stand-Alone— After enabling the front panel Dial, adjust the
Setpoint to cycle the load standard readout between zero and full
tension three times. This removes sensor hysteresis. Path: Status >
Setpoint
Automated— Use the Manual Cmd slider on the Manual
Command window to cycle the load standard readout between
zero and full tension three times. This removes sensor hysteresis.
Example: When calibrating a ±10 kN range, exercise the actuator
between 0 and -10 volts.
3. Apply a tensile force command that is 80% of the full scale range.
Stand-Alone:
A. Using the Dial, adjust the Setpoint for a tensile force
command that is 80% of the full scale range.
B. Verify that your force transducer feedback signal is 80% of the
full scale range.
Automated:
A. Open the Manual Command window, and then click on the
Enable Manual Command.
B. Adjust the Manual Cmd slider on the Manual Command
window for 80% of the full scale range.
C. Use the Station Signals panel to verify that your force
transducer signal equals 80% of the full scale range.

Table of Contents

Other manuals for MTS Systems FlexTest SE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the MTS Systems FlexTest SE and is the answer not in the manual?

MTS Systems FlexTest SE Specifications

General IconGeneral
BrandMTS Systems
ModelFlexTest SE
CategoryController
LanguageEnglish

Related product manuals