Model 493.02 Controller Service
Calibrating a Force Sensor
Calibration
150
12. Record conditioner feedback readings at predetermined tensile
force command points.
Note After shutting down system hydraulics, you will enter these
recorded readings on the Linearization Data window.
Stand-Alone:
A. Use the Dial to adjust the Setpoint for a 0% command.
B. Record the force standard’s readout value and corresponding
conditioner feedback reading in the 0% row of your record
sheet.
C. Use the Dial to adjust the Setpoint for a -2% tensile force
command.
D. Record the force standard’s readout signal and corresponding
conditioner feedback reading in the -2% row of your record
sheet.
E. Repeat steps C and D for other tensile force commands
(typically at -4, -6, -8, -10, -20, -40, -70, and -100 percent of
full scale).
Automated:
A. Adjust the Manual Cmd slider for a 0% command.
B. Record the force standard readout signal and corresponding
conditioner feedback reading at the 0% command line of your
record sheet.
C. Adjust the Manual Cmd slider for a -2% tensile force
command.
D. Record the force standard readout signal and corresponding
conditioner feedback reading at the -2% command line of
your record sheet.
E. Repeat steps C and D for other tensile force commands
(typically at -4, -6, -8, -10, -20, -40, -70, and -100 percent of
full scale).

Table of Contents

Other manuals for MTS Systems FlexTest SE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the MTS Systems FlexTest SE and is the answer not in the manual?

MTS Systems FlexTest SE Specifications

General IconGeneral
BrandMTS Systems
ModelFlexTest SE
CategoryController
LanguageEnglish

Related product manuals