Model 493.02 Controller Service
Calibrating a Force Sensor
Calibration
152
C. Check validity before entering each pair of values on a new
Calibration Data Sheet.
Automated:
A. Adjust the Manual Cmd slider for each retraction and
extension command point on the Linearization Data window.
B. At each command point, verify both the force standard value
(Standard) and its corresponding conditioner feedback value
(Conditioner).
C. Check validity before entering each pair of values on a new
Calibration Data Sheet.