Model 493.02 Controller Service
Calibrating Temposonics Sensors
Calibration
170
For Automated systems, you can also use the Station Signals
panel to monitor sensor output. On the Station Manager Display
menu, select Station Setup. In the navigation pane, select Station
Signals to display the Station Signals panel to monitor current
values for user-defined signals.
For more information on using the Station Signals panel, refer to
“About the Station Signals Panel” in Chapter 3, Station Manager of
the Model 793.00 System Software manual.
Task 2 Create a calibration file
Note This task applies to automated FT SE controllers only.
This task creates a sensor calibration file and sets up any ranges you
may want. A Temposonics sensor does not require ranges, its
resolution is always the same.
See “How to Create a Sensor File” in Chapter 3, Station Manager of the
Model 793.00 System Software manual for a detailed procedure.
Setting Temposonics
resolution
In the Resolution box, enter the Temposonics resolution supplied in
the Temposonics documentation.
Note If want to use units different than those supplied in the sensor
documentation, enter the full scale and resolution in the supplied
units first, and then switch to the desired full-scale units. The units
conversion will be calculated automatically.