Model 493.02 Controller Service
Hydraulic Configurations
225
Independent HSM (shared HPU)
For this multiple controller configuration:
Each controller has an independent hydraulic service manifold
(HSM).
An HPU is shared by all connected controllers via Y-cables.
Select HPU/HSM for Power Type on each connected controller.
Path: Config > Edit Hardware > Power Options > Power Type
> HPU/HSM
FlexTest SE
Emergency Stop
7
8 9
65
4
3
2
1
.
0
+/-
enter
cancel
FG
Setup
Status
Scope
Meters
Tuning
Config
Limits
Log
Menu
?
A
B
Recall
Enabled
reset
Rewind Run
Hold
Stop
Power
Program
Hydraulic
Interlock
Program
Interlock
HPU
HSM
Off
Low
High
Off
Low
High
Navigate
Monitor 1
Monitor 2
FlexTest SE
Emergency Stop
7
8 9
65
4
3
2
1
.
0
+/-
enter
cancel
FG
Setup
Status
Scope
Meters
Tuning
Config
Limits
Log
Menu
?
A
B
Recall
Enabled
reset
Rewind Run
Hold
Stop
Power
Program
Hydraulic
Interlock
Program
Interlock
HPU
HSM
Off
Low
High
Off
Low
High
Navigate
Monitor 1
Monitor 2
Power Type = HPU/HSM
Interlock/Hydraulics
HSM 1
561264-xx
J28 HSM
J25 HPU
397138-xx
563160-xx
P
ower
T
ype =
HPU/HSM
HSM 2
397135-xx
HPU
Converter
HPU
397107-xx

Table of Contents

Other manuals for MTS Systems FlexTest SE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the MTS Systems FlexTest SE and is the answer not in the manual?

MTS Systems FlexTest SE Specifications

General IconGeneral
BrandMTS Systems
ModelFlexTest SE
CategoryController
LanguageEnglish

Related product manuals