10 Time Domain (Scope) Mode
194 IDA-3106 Narda
The following status messages are possible:
To start the recording:
Press the Arm softkey.
The message Armed appears on the screen and the Arm softkey is
now labeled Break.
To break off recording:
Press the Break softkey.
10.5.3 Time Controlled
Time Domain (Scope) • Trigger Menu
I⁄Q Analyzer • Trigger Menu
This trigger type allows you to start the recording at a specified time.
Recording stops as soon as the Time Span has elapsed.
To start the recording immediately:
1. Press the Trigger Mode softkey and select Time Controlled.
The message Waiting for Arm appears on the screen.
2. Press the Start Now softkey.
To start the recording at a specified time:
1. Press the Trigger Mode softkey and select Time Controlled.
The message Waiting for Arm appears on the screen.
2. Press the Start Date softkey and enter the desired date. You can enter
the current date automatically by pressing the Current Date softkey.
Table 36: Trigger messages
Message Function / Explanation
Waiting for Arm Trigger mode has been activated (Single or Multiple) but the
trigger function has not yet been started (Arm softkey)
Waiting This message appears in Multiple trigger mode when the
trigger condition has already been met once and the trigger is
waiting for the next event.
Armed The trigger function has been activated (Arm softkey has
been pressed) but no valid event has occurred yet
Triggered This message appears only briefly when the trigger condition
is met.
Stopped An event has met the trigger condition. The display stops and
the signal trace is "frozen".