6 Common Functions
90 IDA-3106 Narda
6.1 Notes on this chapter
The menu paths shown in gray immediately below the section heading
indicate the operating modes and sub-menus where the function described
can be found.
Screenshots, where used, only serve to illustrate the function. As a rule, the
settings described apply in the same way in the other operating modes
although the display on the screen may not be the same as the screenshot
used in this manual.
6.2 Resetting all device settings
It can be useful to reset the device settings when you have used it for a long
time or it has been used by other persons.
To reset all device settings:
Any menu is open.
1. Press and hold down the ESC key for 3 seconds.
2. Press the OK key.
This resets both the system settings and the measurement settings. The
following are not reset:
• Stored data logger content (measurement results)
• Configuration files (antennas, cables, tables, etc.)
• microSD card contents
• The system settings for: Language, Battery Type, Time Zone, Daylight
Saving Time, Magnetic Declination.
You can reset the system settings (Settings) and measurement settings
(Factory Setup) separately. You can find information about this under
Resetting all device settings on page 74.