10.6 Saving measured values automatically
Narda IDA-3106 195
3. Press the Start Time softkey and enter the desired time. You can enter
the current time automatically by pressing the Current Time softkey.
The start date and time are displayed in the lower status line.
4. Press the Prog Start softkey to activate the trigger function.
The message Armed is displayed on the screen. (If the start time
precedes the actual time, the trigger function starts immediately.)
The recording starts when the specified time has been reached and stops
when the Time Span has elapsed.
To deactivate the trigger function or break off recording:
Press the Break softkey.
10.6 Saving measured values automatically
Time Domain (Scope) • Trigger Menu
I⁄Q Analyzer • Trigger Menu
The Auto Save function allows you to save measured values automatically
when a trigger event occurs.
To save measured values automatically:
1. Open the Trigger menu and select the desired trigger type.
2. Press the Auto Save softkey.
The auto save function is active, the softkey label shows On.
3. Press the Arm softkey.
The trigger function is active; the Suspend Auto Save softkey is
displayed.
The measurement is saved when the trigger event occurs and the
time specified by Time Span has elapsed.
To suspend auto saving:
Press the Suspend Auto Save softkey.