EasyManua.ls Logo

NARDA IDA - Measuring the Duty Cycle and Time Domain Power

NARDA IDA
292 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10 Time Domain (Scope) Mode
196 IDA-3106 Narda
To set the time span between save events for Multiple trigger type:
Auto Save Holdoff specifies the dead time between consecutive save
events. This dead time begins when the measurement time (Time Span)
has elapsed completely following the last trigger event.
Multiple trigger type is selected.
Press the Auto Save Holdoff softkey, select an item using the rotary
control knob, and press OK to set the selection.
To set the maximum number of events to be saved when Multiple
trigger type is set:
The maximum number of events to be saved is set by Auto Save Events.
Multiple trigger type is selected.
Press the Auto Save Events softkey, select an item using the rotary
control knob, and press OK to set the selection.
10.7 Measuring the Duty Cycle and Time
Domain Power
The Duty Cycle evaluation function is available for High Resolution Scope
and Long-Time Scope modes. This function automatically displays the ratio
of the average to the maximum power level (Pavg/Ppeak), calculated for the
displayed time period. The Duty Cycle function corresponds to the signal
duty cycle if the signal shape is rectangular. The values of Pavg and Ppeak
are also displayed in the Marker bar. The measured value for Pavg is often
referred to as the Time Domain Power.

Table of Contents

Related product manuals