17 Specifications
252 IDA-3106 Narda
17.3 Operating modes
17.3.1 Overview
EMC • Complies with EMC Directive 2014/30/EU (previously
2004/108/EC) and EN 61326-1: 2013
• Immunity: IEC/EN: 61000-4-2, 61000-4-3, 61000-4-4,
61000-4-5, 61000-4-6, 61000-4-11
Immunity at least 200 V/m
(RF input power limited to permissible values)
• Emissions: IEC/EN: 61000-3-2, 61000-3-3,
IEC/EN 55011 (CISPR 11) Class B
Ingress protection • IP 52 (with antenna attached and interface protector closed)
• IP 67 (stored in the hardcase)
Safety Complies with European Low Voltage Directive 2014/35/EU
(previously 2006/95/EC) and EN 61010-1: 2010
Air humidity
(operating range) RF
< 29 g/m
3
(< 93 % RH at +30 °C)
Weight 2.8 kg / 6.2 lbs (basic unit including rechargeable battery)
Dimensions (H x W x D) 213 mm x 297 mm x 77 mm (8.4" x 11.7" x 3.0")
Power supply Rechargeable battery
• Lithium-Ion rechargeable battery pack,
“hot-swappable” during operation
• operating time: 3 hours (nominal)
• charging time: 5.5 hours (nominal)
External power supply
• Input: 9 to 15 VDC
• Adapter: 100 – 240 VAC / 12 V DC, 2.5 A
Recommended calibration
interval
24 months
Country of origin Germany
Measurements vs.
frequency
• Spectrum Analysis (including Spectrogram)
• Multi-Channel Power
Measurements vs.
time
• Level Meter
• Time Domain (Scope)
• I⁄Q Analyzer
Measurement vs.
orientation/position
• Direction Finding
including Horizontal Scan and Localization