114/118
Alarm
Action level 1
Action level 2
Action level 3
First drip of
condensate false
detection :
Condenser not cleaned up or water condensation
condensation may affect the photo
detectors
contact technical services
Level follower
fork problem :
First drip fork
problem :
Heating element
hot :
Let heating element cool down until fan stops
check if the fan is working
correctly
contact technical services
Condenser bath
not clean :
IBP off-standard :
ill-adapted program file, heating plate or sample
5% recovered off-
standard :
ill-adapted program file, heating plate or sample
10% recovered
off-standard :
ill-adapted program file, heating plate or sample
FBP off-standard
:
ill-adapted program file, heating plate or sample
Condenser bath
control problem :
condenser temperature out of range (<-5°C or >70°C
) check condenser temperature
contact technical services