N O R M A L A B F r a n c e
Receiver
chamber control
problem :
condenser temperature out of range (<-5°C or >70°C
) check condenser temperature
contact technical services
check fuse of heating element
contact technical services
Power supply
problem for X
second(s) :
check power supply stability (network)
contact technical services
Power supply
problem upper 5
minutes
:
check power supply stability (network)
contact technical services
check heating compartment before starting a new
test
Barometric
sensor problem :
check the pressure sensor (back of the instrument)
switch to manual barometric
pressure correction
contact technical services
check Valve position (nitrogen bottle)
Bottle valve open and manometer
at 0, empty bottle
Refill the bottle with nitrogen
Flask heater area
hostile fire
check heating block and tray
check Thermocouple (above the
tray)
contact technical services