Depending on the equipment available, there are two possibilities:
- one or two baths stabilised at temperature are available and it is possible to carry out
calibration with the probe.
- One Pt100 probe simulator (decade resistor or calibrator): the probe will be replaced by
the simulator.
Calibration by bath:
- Disconnect the probe and insert the extension (accessory reference 60525) between the
probe and the instrument so the probe can be immersed in each calibration bath.
- Select the calibration mode (Real).