The instrument waits for the value read off to
stabilise (value stable at 800 conversion points) and
calculates the average of five acquisitions.
Confirm by selecting the green tick, restart
measuring by selecting ‘restart’ mode or cancel and
exit by selecting the red cross.
NB: to make a calibration with sufficient
accuracy, the system imposes a
difference of 80°C between the two
calibration points.
If preselections are not appropriate, reply ‘yes’
and use the keyboard on screen to enter the
new point values. End each input session with
a green tick.
The procedure requires placing the probe in
the lowest temperature bath (example
50.3°C) and confirming by ‘OK’.