63/118
The procedure displays the calibration
points with the conversion points and
asks for confirmation or otherwise.
If confirmation is accepted, calibration
will be displayed and can be used when
selected by default.
NB: If the signal is too unstable, the instrument will not achieve stability and remains in this function.
It is possible to force it by pressing the green tick (but the calibration will not be accurate given the
instability).
Follow the same procedure for the high
calibration point, place the probe in the warmest
bath and confirm with the green tick.
Calibration with a simulator:
Calibration by simulation with a Pt 100 calibrator (or decade box) is similar to calibration by bath. It
requires the calibration cable reference 60522 to connect the calibrator in place of the probe.
This calibration allows calibrating the acquisition system of the instrument, excluding the
temperature probe. Error introduced by this sensor is not incorporated.