EasyManua.ls Logo

OCULUS Corvis ST - Page 13

OCULUS Corvis ST
86 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4 Safety Instructions
Instruction Manual Corvis® ST (G/72100/XXXX/EN 1019 Rev01) 7 / 80
In the patient environment, use devices that conform to IEC 60601-1. If
a multiple power socket is to be used, or if a device is to be used that does
not meet the IEC 60601-1 standard, use an isolating transformer.
Information about the operation of an ME system
The Corvis® ST and a connected computer form a medical
electrical system (ME system) according to DIN EN 60601-1. If you
connect additional devices, such as, for example a printer, those devices
become part of the ME system.
Make sure that all devices of the ME system meet the requirements
of IEC 60601-1 or IEC 60950-1.
Note that an output voltage of maximum 5.5 V DC is supplied by a
device connected via USB.
Abb. 4-1: Patient environment

Table of Contents