21 Appendix
70 / 80 Instruction Manual Corvis® ST (G/72100/XXXX/EN 1019 Rev01)
Warning
The use of accessories, transducers and cables not specified by OCULUS
(for example as replacement parts) may result in increased emissions or
decreased immunity of the Corvis® ST.
Use only the original accessories, transducers and cables specified by
OCULUS.
The use of accessories, transducers and cables specified by OCULUS with
devices other then the Corvis® ST may result in increased emissions or
decreased immunity of the other device.
Do not use the accessories, transducers and cables specified by
OCULUS with devices other then the Corvis® ST.
To be in compliance with the requirements of the IEC 60601-1-2 6.1 and
6.2 the following types of equipment, accessories, power adapters and ca-
bles must be used.
Order number Description
72100
72200
Corvis® ST
05200905 Cable with plug, EU standard 1.8 m
05200910 Cable with plug, US standard 1.8 m
05200915 Cable with plug, GB standard 1.8 m
05200920 Cable with plug, AU standard 1.8 m
05200601 USB cable
015692000010 USB FS MED Isolator
05150285 Power Supply 15 V DC; 6 A