Cal Test
This feature checks the calibration against last stored calibration information. The Cal Error
display indicates the difference since the last automatic calibration.
1. Make sure there is no load on the pan and close the chamber doors.
2.
ON
TARE
.
3.
ON
TARE
4.
MODE
.
5.
ON
TARE
.
Cal End
To exit the calibration menus, proceed as follows:
1.
ON
TARE
.
2.
ON
TARE
.
3.
MODE
.
4.
ON
TARE
.
5.
MODE
.
6.
ON
TARE
.
gs
^]emu cal
auTo
TesT
CAL
CAL
-c-
-c-
0.00000
.
g
g
CAL
ERROR
0.00000
gs
^]emu cal
auTo
CAL
eND
CAL
cal
MENU
eND
0.00000