64
MAINTENANCE
MODEL AP110 AP210 AP310 AP250D
Capacity (g) 110 210 310 52/210
Readability (mg) 0.1 0.01/0.1
Weighing modes g, oz, ct, dwt, (3) taels, oz t, gn, mommes,
1 custom unit, parts counting, percent weighing,
Repeatability (Std. dev.) (mg) 0.1 0.2 0.02/0.1
Linearity (mg) ±0.2 ±0.5 ±0.03/0.2
Tare range To capacity by subtraction
Stabilization time (sec) 4 8 12/5
Sensitivity drift (10 - 30 °C) ±2 ppm/ °C
Operating temperature range 50° to 104°F/10° to 40°C
Relative humidity max. 80 %
Installation category II
Pollution degree 2
Calibration Motorized internal and manual external
Internal calibration weights Stainless steel - measured to ±0.1 mg at an
air density of 1.2 mg/cm
3
on virtual mass with density 8.0 g/cm
3
Power requirements 100-120 V, 110 mA, 50/60Hz
220-240 V, 55 mA, 50/60Hz
Display (in/cm) Vacuum fluorescent (0.5/1.3 high)
Pan size (in/cm) 3.5/9
Free height above pan (in/cm) 9.5/24
Dimensions 7.7 x 12.2 x 15.7
(WxHxD) (in/cm) 19.5 x 31 x 40
Net weight (lb/kg) 22.5/10.2
SPECIFICATIONS
Menu
Menu

Table of Contents