62
MAINTENANCE
Error Codes List (Cont.)
System Errors
5.1 thrrough 5.9
System errors. Turn the power off, then turn it back on. If error persists, have the balance
serviced.
Checksum Errors
9.7 Invalid setup data checksum. Check Setup, User, and Print menus settings. If
possible, try to enter menus and exit using END to restore menu settings. May
be caused by a faulty component, or in rare cases, a severe static charge. If
error persists, balance must be serviced.
If your balance displays ERR 9.7 briefly at switch-on, please follow the
procedure below:
11.
ON
TARE
12.
PRINT
13.
ON
TARE
14.
MODE
15.
ON
TARE
16.
ON
TARE
17.
ON
TARE
18.
PRINT
19.
ON
TARE
20.
MODE
21.
ON
TARE
Switch balance on:
1.
ON
TARE
2.
MODE
3.
ON
TARE
4.
ON
TARE
5.
ON
TARE
6.
PRINT
7.
ON
TARE
8.
MODE
9.
ON
TARE
10.
ON
TARE
^]emu
useR
ReseT
yes
ReseT
eND usR
useR
seTup
ReseT
yes
ReseT
eND
seTup
pR INT
ReseT
yes
ReseT
eND
pR INT
eND