98
If the phenomena are not improved even after you took the above remedial action, contact Olympus.
Please provide us the following information at that time. If this product is combined with PC, please provide us the PC
model name, OS, free space in HDD, memory, and CPU.
· Product name and abbreviated name (example: Control box D2-CB)
· Product number
· Problem: If the error code (4-digit number) is displayed, please inform us as well.
The repair parts are kept for the period of five years after purchase.
Repair request
Problem Cause Remedy Page
Functions linked with
microscope are not working.
Cables are not connected properly. Connect cables of the microscope,
control box for microscope (BX3-CBH,
etc.) and control box D2-CB properly.
90
The control box (BX3-CBH, etc.) is not
recognized.
Turn OFF the main switches of the
control box for microscope (BX3-CBH,
etc.) and control box D2-CB. Then,
turn ON in the order of
1
Control
box for microscope (BX3-CBH, etc.)
and
2
Control box D2-CB.
92
The automatic scale change
function does not work.
The objective magnification is not
specified. Or it is not set properly.
Specify the magnification of the
objective you are using.
92
When the main switch of the
control box D2-CB is turned
OFF, various settings or the
scale function are returned to
the status before setting.
When turning OFF the main switch, the
system is terminated forcibly (the main
switch is long-pressed.)
When turning OFF the main switch, do
not long-press the main switch. You
can turn OFF the power by pressing
the main switch lightly.