EasyManua.ls Logo

Olympus DP23-AOU - Connecting Procedures

Olympus DP23-AOU
114 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
96
8-2 Connecting procedures
For procedures to connect the microscope to the control box, refer to the instruction manual of the relevant control box
or the microscope.
· Remove only the covers or stickers of necessary connectors on the control box.
· Be sure to connect only units specified by Olympus to each connector. If units which are not specified
by Olympus are connected, the full performance is not ensured.
· Use the RS232C cable for connecting DP2-AOU to the control box for microscope.
· Be sure to set the main switch of each unit to OFF before connecting the cables.
· Align the orientation of the connector and insert it completely to the end. If the connector has clamping
screws, be sure to secure them.
· Inserting the connectors which are not specified by Olympus may damage them.
The connecting procedures are described below by using U-CBS as an example. For other connecting procedures, refer
to instruction manuals of devices to be connected.
NOTE
Connecting to U-CBS
U-CBS
Connector of the coded illuminator
DP2-AOU
RS232C connector
AC adapter for U-CBS
When attaching the connector of the coded revolving nosepiece or the coded
illuminator, insert it by aligning
marks
A
.
U-HSEXP connector
RS232C cable
(provided with U-CBS)
A
Connector of the coded
revolving nosepiece

Table of Contents

Related product manuals