12.2 Scanner Interface Connector
Description
DE-15, female connector
Manufacturer, number
Kycon, K61-E15S-NS
Suggested cable connector
Conec, 301A10129X
Figure 12-2 The scanner interface connector
Table 24 Pinout for the scanner interface connector
Pin I/O Signal Description Current Level
1In DIN1/
Preset1
Digital input 1/preset
axis 1.
Programmable input. Can
be configured as generic
input 1, or as the preset of
encoder 1. Refer to the
OmniScan software users
manuals (“Configuring the
Digital Input”) for
instructions on
programming this input.
To preset, you must use a
high-level signal with a
minimum signal length of
50 ms.
TTL
1
5
1
0
11
15
6