OMNI-M-PA16128 Module Specifications
155
G.2 Acquisition Specifications
This section lists the acquisition specifications for frequency, data display, and
synchronization (see Table 47 on page 155).
G.3 Data Specifications
This section details the data specifications of processing, TCG, and storage (see
Table 48 on page 155).
Table 47 Acquisition specifications — OMNI-M-PA16128
Frequency
Digitizing frequency 100 MHz (10 bits)
A-scan (acquisition
mode)
Up to 6000 A-scans per second (512-point,
8-bit A-scan)
Maximum pulsing rate Up to 10 kHz (C-scan)
Depth in material 59.8 meters in steel (longitudinal wave),
10 ms with compression
0.49 meter in steel (longitudinal wave),
81.9 µs without compression
Display
Refresh rate A-scan: 60 Hz; S-scan: 20 Hz to 30 Hz
Envelope (echo-dynamic
mode)
Yes: Volume-corrected S-scan (30 Hz)
Synchronization
On internal clock 1 Hz to 10 kHz
External pace Yes
On encoder On 2 axes: from 1 to 65536 steps
Table 48 Data specifications — OMNI-M-PA16128
Processing
Number of data points Up to 8000
Real-time averaging 2, 4, 8, and 16
Rectification RF, full wave, half wave+, half wave