12.3 Alarm and I/O Connector
Description
DE-9, female connector
Manufacturer, number
Amphenol, 788797-1
Suggested cable connector
ITT Cannon, DE-9P
Figure 12-3 The alarm and I/O connector
Table 25 Pinout for the alarm and I/O connector
Pin I/O Signal Description Current Level
1 Out Al1 Alarm output 1. Disabled
on reset, and is at 0 V. When
active, it is at 5 V.
±15 mA TTL
2 Out Al2 Alarm output 2. Disabled
on reset, and is at 0 V. When
active, it is at 5 V.
±15 mA TTL
3 Out Al3 Alarm output 3. Disabled
on reset, and is at 0 V. When
active, it is at 5 V.
±15 mA TTL
4 Out AOUT1 Analog output 1 ±15 mA ±5 V
5 Out AOUT2 Analog output 2 ±15 mA ±5 V
6– GNDGround