OMNI-M2-PA1664/16128/32128/32128PR Module Specifications
191
When the UT channel is used in pulse-echo mode, pulses occur on both the P1 and R1
connectors. When pulse-echo mode is selected, Olympus recommends using only the
P1 connector to connect the probe.
Receiver
Gain range 0 dB to 120 dB maximum input signal
34.5 Vp-p (full-screen height)
Input impedance 60 Ω in pulse-echo mode
50 Ω in pulse-receive mode
System bandwidth 0.25 MHz to 28 MHz (−3dB)
Table 70 Acoustic specifications — PA channel
Pulser
Voltage 40 V, 80 V, and 115 V
Pulse width Adjustable from 30 ns to 500 ns; resolution of 2.5 ns
Fall time <10 ns
Pulse shape Negative square pulse
Output impedance For 32128PR model:
35 Ω in pulse-echo
30 Ω in pitch-catch
For other models:
25 Ω
Receiver
Gain range 0 dB to 80 dB maximum input signal 550 mVp-p (full-
screen height).
Input impedance For 32128PR model:
50 Ω in pulse-echo
90 Ω in pitch-catch
For other models:
65 Ω
System bandwidth 0.6 MHz to 18 MHz (−3dB)
Beam formation