Olympus OmniScan MX - Table 70 Acoustic Specifications - PA Channel

Olympus OmniScan MX
230 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DMTA-20015-01EN [U8778402], Rev. M, February 2017
OMNI-M2-PA1664/16128/32128/32128PR Module Specifications
191
When the UT channel is used in pulse-echo mode, pulses occur on both the P1 and R1
connectors. When pulse-echo mode is selected, Olympus recommends using only the
P1 connector to connect the probe.
Receiver
Gain range 0 dB to 120 dB maximum input signal
34.5 Vp-p (full-screen height)
Input impedance 60 in pulse-echo mode
50 in pulse-receive mode
System bandwidth 0.25 MHz to 28 MHz (3dB)
Table 70 Acoustic specifications — PA channel
Pulser
Voltage 40 V, 80 V, and 115 V
Pulse width Adjustable from 30 ns to 500 ns; resolution of 2.5 ns
Fall time <10 ns
Pulse shape Negative square pulse
Output impedance For 32128PR model:
35 in pulse-echo
30 in pitch-catch
For other models:
25
Receiver
Gain range 0 dB to 80 dB maximum input signal 550 mVp-p (full-
screen height).
Input impedance For 32128PR model:
50 in pulse-echo
90 in pitch-catch
For other models:
65
System bandwidth 0.6 MHz to 18 MHz (3dB)
Beam formation
Table 69 Acoustic specifications — UT channel using UT connector (continued)
NOTE